Medtronic MyCareLink Patient Monitor

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 7.0
  • ATTENTION: Low attack complexity
  • Vendor: Medtronic
  • Equipment: MyCareLink Patient Monitor 24950, 24952
  • Vulnerabilities: Cleartext Storage of Sensitive Information, Empty Password in Configuration File, Deserialization of Untrusted Data

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could lead to system compromise, unauthorized access to sensitive data, and manipulation of the monitor’s functionality.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Medtronic products are affected:

  • MyCareLink Patient Monitor model 24950: All versions
  • MyCareLink Patient Monitor model 24952: All versions

3.2 Vulnerability Overview

3.2.1 CLEARTEXT STORAGE OF SENSITIVE INFORMATION CWE-312

Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files.

CVE-2025-4394 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4394. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 EMPTY PASSWORD IN CONFIGURATION FILE CWE-258

Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access/modify system functionality.

CVE-2025-4395 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4395. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.3 DESERIALIZATION OF UNTRUSTED DATA CWE-502

Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges.

CVE-2025-4393 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.5 has been calculated; the CVSS vector string is (AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4393. A base score of 5.9 has been calculated; the CVSS vector string is (AV:L/AC:H/AT:P/PR:L/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Ireland

3.4 RESEARCHER

Ethan Morchy from Somerset Recon and Carl Mann, an independent researcher, reported these vulnerabilities to Medtronic.

4. MITIGATIONS

The identified vulnerabilities were reported as low-risk findings. An attacker would need to physically tamper with the monitor to exploit them. In response, starting in June 2025, Medtronic began deploying security updates to address these findings.

Medtronic recommends the following actions:

  • The security update process is performed automatically when the monitor is connected to the internet. Users should ensure that their remote monitor is plugged in to receive updates.
  • Physicians should continue to prescribe monitors as intended.
  • Users should maintain possession of their home monitor.
  • Users should only use home monitors provided directly from a healthcare provider or a Medtronic representative.

Users needing additional assistance should contact security@medtronic.com.

For more information regarding these vulnerabilities, refer to Medtronic’s security bulletin.

Users should follow CISA’s guidance in the following areas:

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.

5. UPDATE HISTORY

  • July 24, 2025: Initial Publication

Mitsubishi Electric CNC Series

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v3 7.0
  • ATTENTION: Exploitable from a local network
  • Vendor: Mitsubishi Electric
  • Equipment: CNC Series
  • Vulnerability: Uncontrolled Search Path Element

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to execute malicious code by getting setup-launcher to load a malicious DLL.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of Mitsubishi Electric CNC Series are affected:

  • NC Designer2: All versions
  • NC Designer: All versions
  • NC Configurator2: All versions
  • NC Analyzer2: All versions
  • NC Analyzer: All versions
  • NC Explorer: All versions
  • NC Monitor2: All versions
  • NC Monitor: All versions
  • NC Trainer2: “AB” and prior
  • NC Trainer2 plus: “AB” and prior
  • NC Trainer: All versions
  • NC Trainer plus: All versions
  • NC Visualizer: All versions
  • Remote Monitor Tool: All versions
  • MS Configurator: All versions
  • Mitsubishi Electric Numerical Control Device Communication Software (FCSB1224): All versions
  • Mitsubishi Electric CNC communication software runtime library M70LC/M730LC: All versions
  • NC Virtual Simulator: All versions

3.2 Vulnerability Overview

3.2.1 UNCONTROLLED SEARCH PATH ELEMENT CWE-427

Malicious code execution vulnerability via DLL hijacking due to Uncontrolled Search Path Element (CWE-427) exists in Flexera InstallShield used in multiple software tools and industrial IoT-related products for Mitsubishi Electric CNC Series.

CVE-2016-2542 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.0 has been calculated; the CVSS vector string is (AV:L/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Critical Manufacturing
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Japan

3.4 RESEARCHER

Sahil Shah reported this vulnerability to Mitsubishi Electric.

4. MITIGATIONS

The vulnerability is fixed in the following products and versions.  Download and install the fixed version from the Mitsubishi Electric download site:

  • NC Trainer2: “AC” or later
  • NC Trainer2 plus: “AC” or later

Note that there are no plans to release fixed versions for the following products:

  • NC Designer
  • NC Analyzer
  • NC Monitor
  • NC Trainer / NC Trainer plus
  • NC Visualizer
  • Remote Monitor Tool
  • MS Configurator

For users of products that do not have a fixed version or who cannot immediately update the product, Mitsubishi Electric recommends taking the following mitigations to minimize the risk of exploiting this vulnerability:

  • Restrict physical access to the computer using the product.
  • Install an antivirus software in the computer using the affected product.
  • Do not open untrusted files or click untrusted links.
  • Do not run setup-launchers obtained from sources other than our branches, distributors or the Mitsubishi Electric FA website.
  • Before running the setup-launcher, make sure that no DLL exists in the folder containing the setup-launcher executable file (the name varies depending on the product) for the product.

For more information, see Mitsubishi Electric 2025-008.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely. This vulnerability has a high attack complexity.

5. UPDATE HISTORY

  • July 24, 2025: Initial Republication of Mitsubishi Electric 2025-008

Schneider Electric EcoStruxure Power Operation

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v3 8.8
  • ATTENTION: Exploitable remotely/low attack complexity/public exploits are available/known public exploitation
  • Vendor: Schneider Electric
  • Equipment: EcoStruxure Power Operation
  • Vulnerabilities: Improper Neutralization of Directives in Dynamically Evaluated Code (‘Eval Injection’), Integer Overflow to Buffer Overflow, Improper Handling of Highly Compressed Data (Data Amplification), Out-of-bounds Write, Uncontrolled Resource Consumption

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could result in the loss of system functionality or unauthorized access to system functions.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

Schneider Electric reports the following products use an affected version of the PostgreSQL database server:

  • EcoStruxure Power Operation (EPO): 2022 CU6 and prior
  • EcoStruxure Power Operation (EPO): 2024 CU1 and prior

3.2 Vulnerability Overview

3.2.1 Improper Neutralization of Directives in Dynamically Evaluated Code (‘Eval Injection’) CWE-95

Pillow Version 10.1.0 allows PIL.ImageMath.eval arbitrary code execution via the environment parameter. This is a different vulnerability from CVE-2022-22817, which pertains to the expression parameter.

CVE-2023-50447 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.1 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H).

3.2.2 Integer Overflow to Buffer Overflow CWE-680

In _imagingcms.c in Pillow prior to 10.3.0, a buffer overflow exists because strcpy is used instead of strncpy.

CVE-2024-28219 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.7 has been calculated; the CVSS vector string is (CVSS:3.1/AV:L/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H).

3.2.3 Improper Handling of Highly Compressed Data (Data Amplification) CWE-409

Versions of Pillow before 9.2.0 improperly handle highly compressed GIF data (data amplification).

CVE-2022-45198 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

3.2.4 Out-of-bounds Write CWE-787

A heap buffer overflow in vp8 encoding in libvpx, used by Google Chrome versions prior to 117.0.5938.132 and libvpx Version 1.13.1 allowed a remote attacker to potentially exploit heap corruption via a crafted HTML page.

CVE-2023-5217 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.8 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

3.2.5 Uncontrolled Resource Consumption CWE-400

Envoy is a cloud-native high-performance edge/middle/service proxy. Envoy’s HTTP/2 codec may leak a header map and bookkeeping structures upon receiving RST_STREAM  immediately followed by the GOAWAY frames from an upstream server. In nghttp2, cleanup of pending requests due to receipt of the GOAWAY frame skips de-allocation of the bookkeeping structure and pending compressed header. The error return [code path] is taken if the connection is already marked for not sending more requests due to GOAWAY frame. The clean-up code is right after the return statement, causing a memory leak. This results in denial of service through memory exhaustion. This vulnerability was patched in Versions 1.26.3, 1.25.8, 1.24.9, 1.23.11.

CVE-2023-35945 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

3.2.6 Uncontrolled Resource Consumption CWE-400

The HTTP/2 protocol allows a denial of service (server resource consumption) because request cancellation can reset many streams quickly, as was exploited in the wild from August to October 2023.

CVE-2023-44487 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Commercial Facilities, Critical Manufacturing, Energy
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: France

3.4 RESEARCHER

Schneider Electric reported these vulnerabilities to CISA.

4. MITIGATIONS

EcoStruxure Power Operation 2024 CU2 includes fixes for these vulnerabilities and is available for download.

Schneider Electric recommends users to employ appropriate patching methodologies when applying these patches to their systems. They strongly recommend making backups and evaluating the impact of these patches in a test and development environment or on offline infrastructure. Contact Schneider Electric’s Customer Care Center for assistance removing a patch.

If users choose not to apply the remediation mentioned above, Schneider Electric recommends the following:

  • If waveform analysis and ETAP simulation features are not used, uninstall PostgreSQL,
    OR
  • For users of waveform analysis and ETAP simulation features, Schneider Electric recommends all deployments of EPO only accept connections from localhost in PostgresSQL. Contact Schneider Electric’s Customer Care Center for information on how to modify PostgreSQL. Additionally, Schneider Electric recommends users manually uninstall PostgreSQL 14.10 and update to PostgreSQL 14.17 or higher.

For more information, see the associated Schneider Electric security advisory SEVD-2025-189-03: EcoStruxure Power Operation PDF version, CSAF version.

Schneider Electric strongly recommends adhering to the following industry cybersecurity best practices:

  • Locate control and safety system networks and remote devices behind firewalls, and isolate them from the business network.
  • Install physical controls to prevent unauthorized personnel from accessing industrial control and safety systems, components, peripheral equipment, and networks.
  • Place all controllers in locked cabinets and never leave them in the “Program” mode.
  • Never connect programming software to any network other than the one intended for that device.
  • Scan all methods of mobile data exchange with the isolated network, such as CDs, USB drives, etc., before use in terminals or any nodes connected to these networks.
  • Never allow mobile devices that have connected to any network other than the intended network to connect to safety or control networks without proper sanitation.
  • Minimize network exposure for all control system devices and systems, ensuring they are not accessible from the Internet.
  • When remote access is required, use secure methods such as virtual private networks (VPNs). Recognize that VPNs may have vulnerabilities and should be updated to the most current version available. Also, understand that VPNs are only as secure as the connected devices.
    For more information, refer to the Schneider Electric recommended cybersecurity best practices document.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

5. UPDATE HISTORY

  • July 22, 2025: Initial Republication of Schneider Electric SEVD-2025-189-03

Schneider Electric EcoStruxture IT Data Center Expert

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 9.5
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Schneider Electric
  • Equipment: EcoStruxure IT Data Center Expert
  • Vulnerabilities: Improper Neutralization of Special Elements used in an OS Command (‘OS Command Injection’), Insufficient Entropy, Improper Control of Generation of Code (‘Code Injection’), Server-Side Request Forgery (SSRF), Improper Privilege Management, and Improper Restriction of XML External Entity Reference

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to disrupt operations and access system data.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

Schneider Electric reports the following product is affected:

  • EcoStruxure IT Data Center Expert: Versions v8.3 and prior

3.2 Vulnerability Overview

3.2.1 IMPROPER NEUTRALIZATION OF SPECIAL ELEMENTS USED IN AN OS COMMAND (‘OS COMMAND INJECTION’) CWE-78

An improper neutralization of special elements used in an OS command (‘OS command injection’) vulnerability exists, which could cause unauthenticated remote code execution when a malicious folder is created via the HTTP web interface when enabled. HTTP is disabled by default.

CVE-2025-50121 has been assigned to this vulnerability. A CVSS v3.1 base score of 10.0 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-50121. A base score of 9.5 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:L/SA:H).

3.2.2 INSUFFICIENT ENTROPY CWE-331

An insufficient entropy vulnerability exists, which could cause the root password discovery when the password generation algorithm is reverse engineered with access to installation or upgrade artifacts.

CVE-2025-50122 has been assigned to this vulnerability. A CVSS v3.1 base score of 8.3 has been calculated; the CVSS vector string is (CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-50122. A base score of 8.9 has been calculated; the CVSS vector string is (CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:L/SA:H).

3.2.3 IMPROPER CONTROL OF GENERATION OF CODE (‘CODE INJECTION’) CWE-94

An improper control of generation of code (‘code injection’) vulnerability exists, which could cause remote command execution by a privileged account when the server is accessed via a console and the hostname input is exploited.

CVE-2025-50123 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.2 has been calculated; the CVSS vector string is (CVSS:3.1/AV:P/AC:L/PR:H/UI:N/S:C/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-50123. A base score of 7.2 has been calculated; the CVSS vector string is (CVSS:4.0/AV:P/AC:L/AT:P/PR:H/UI:N/VC:H/VI:H/VA:H/SC:H/SI:L/SA:H).

3.2.4 SERVER-SIDE REQUEST FORGERY (SSRF) CWE-918

A server-side request forgery (SSRF) vulnerability exists, which could cause unauthenticated remote code execution when the server is accessed via the network with knowledge of hidden URLs and manipulation of the host request header.

CVE-2025-50125 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.2 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-50125. A base score of 6.3 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:L/VI:L/VA:N/SC:L/SI:N/SA:N).

3.2.5 IMPROPER PRIVILEGE MANAGEMENT CWE-269

An improper privilege management vulnerability exists, which could cause privilege escalation when the server is accessed by a privileged account via a console and through exploitation of a setup script.

CVE-2025-50124 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.9 has been calculated; the CVSS vector string is (CVSS:3.1/AV:P/AC:H/PR:H/UI:N/S:C/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-50124. A base score of 7.2 has been calculated; the CVSS vector string is (CVSS:4.0/AV:P/AC:H/AT:P/PR:H/UI:N/VC:H/VI:H/VA:H/SC:H/SI:L/SA:H).

3.2.6 IMPROPER RESTRICTION OF XML EXTERNAL ENTITY REFERENCE CWE-611

An improper restriction of XML external entity reference vulnerability exists, which could cause manipulation of SOAP API calls and XML external entities injection, resulting in unauthorized file access when the server is accessed via the network using an application account.

CVE-2025-6438 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:C/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-6438. A base score of 5.9 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:P/PR:H/UI:N/VC:H/VI:N/VA:N/SC:L/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Critical Manufacturing
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: France

3.4 RESEARCHER

Jaggar Henry and Jim Becher of KoreLogic, Inc. reported these vulnerabilities to Schneider Electric.

4. MITIGATIONS

Schneider Electric has identified the following specific workarounds and mitigations users can apply to reduce risk:

  • Schneider Electric EcoStruxure IT Data Center Expert Version 8.3 and prior: Version 9.0 of EcoStruxure IT Data Center Expert includes fixes for these vulnerabilities and is available upon request from Schneider Electric’s Customer Care Center.
  • If users choose not to apply the remediation provided above, they should immediately apply the following mitigations to reduce the risk of exploit:
    • Harden the DCE instance according to the cybersecurity best practices documented in the EcoStruxure IT Data Center Expert Security Handbook

For more information see the associated Schneider Electric CPCERT security advisory SEVD-2025-189-01 EcoStruxure IT Data Center Expert – SEVD-2025-189-01 PDF Version, EcoStruxure IT Data Center Expert – SEVD-2025-189-01 CSAF Version.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs). Recognize VPNs may have vulnerabilities, should be updated to the most recent version available, and are only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • July 22, 2025: Initial Republication of Schneider Electric CPCERT SEVD-2025-189-01

DuraComm DP-10iN-100-MU

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: DuraComm Corporation
  • Equipment: SPM-500 DP-10iN-100-MU
  • Vulnerabilities: Cleartext Transmission of Sensitive Information, Missing Authentication for a Critical Function, Improper Neutralization of Input During Web Page Generation

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to disclose sensitive information or cause a denial-of-service condition.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of DuraComm SPM-500 DP-10iN-100-MU, a power distribution panel, are affected:

  • SPM-500 DP-10iN-100-MU: Version 4.10 and prior

3.2 Vulnerability Overview

3.2.1 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION (‘CROSS-SITE SCRIPTING’) CWE-79

The affected product is vulnerable to a cross-site scripting (XSS) attack. This could allow an attacker to prevent legitimate users from accessing the web interface.

CVE-2025-41425 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.1 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-41425. A base score of 7.2 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 MISSING AUTHENTICATION FOR CRITICAL FUNCTION CWE-306

The affected product lacks access controls for a function that should require user authentication. This could allow an attacker to repeatedly reboot the device.

CVE-2025-48733 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

A CVSS v4 score has also been calculated for CVE-2025-48733. A base score of 8.7 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N).

3.2.3 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

The affected product transmits sensitive data without encryption over a channel that could be intercepted by attackers.

CVE-2025-53703 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-53703. A base score of 8.7 has been calculated; the CVSS vector string is (CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Energy
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Brandon Vincent of Arizona Public Service reported these vulnerabilities to CISA.

4. MITIGATIONS

DuraComm recommends users update to Version 4.10A. Contact DuraComm to obtain the update.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • July 22, 2025: Initial Publication.

CISA Adds One Known Exploited Vulnerability, CVE-2025-53770 “ToolShell,” to Catalog

CISA has added one new vulnerability to its Known Exploited Vulnerabilities Catalog, based on evidence of active exploitation. See CISA’s Alert Microsoft Releases Guidance on Exploitation of SharePoint Vulnerability (CVE-2025-53770) for more information and to apply the recommended mitigations. 

  • CVE-2025-53770: Microsoft SharePoint Server Remote Code Execution Vulnerability

These types of vulnerabilities are frequent attack vectors for malicious cyber actors and pose significant risks to the federal enterprise.

Binding Operational Directive (BOD) 22-01: Reducing the Significant Risk of Known Exploited Vulnerabilities established the Known Exploited Vulnerabilities Catalog as a living list of known Common Vulnerabilities and Exposures (CVEs) that carry significant risk to the federal enterprise. BOD 22-01 requires Federal Civilian Executive Branch (FCEB) agencies to remediate identified vulnerabilities by the due date to protect FCEB networks against active threats. See the BOD 22-01 Fact Sheet for more information.

Although BOD 22-01 only applies to FCEB agencies, CISA strongly urges all organizations to reduce their exposure to cyberattacks by prioritizing timely remediation of Catalog vulnerabilities as part of their vulnerability management practice. CISA will continue to add vulnerabilities to the catalog that meet the specified criteria.

Microsoft Releases Guidance on Exploitation of SharePoint Vulnerability (CVE-2025-53770)

CISA is aware of active exploitation of a new remote code execution (RCE) vulnerability enabling unauthorized access to on-premise SharePoint servers. While the scope and impact continue to be assessed, the new Common Vulnerabilities and Exposures (CVE), CVE-2025-53770, is a variant of the existing vulnerability CVE-2025-49706 and poses a risk to organizations. This exploitation activity, publicly reported as “ToolShell,” provides unauthenticated access to systems and enables malicious actors to fully access SharePoint content, including file systems and internal configurations, and execute code over the network. 

CISA recommends the following actions to reduce the risks associated with the RCE compromise: 

  • Configure Antimalware Scan Interface (AMSI) in SharePoint and deploy Microsoft Defender AV on all SharePoint servers.
    • If AMSI cannot be enabled, disconnect affected products from service that are public-facing on the internet until official mitigations are available. Once mitigations are provided, apply them according to CISA and vendor instructions.
    • Follow the applicable BOD 22-01 guidance for cloud services or discontinue use of the product if mitigations are not available.
  • For information on detection, prevention, and advanced threat hunting measures, see Microsoft’s Customer Guidance for SharePoint Vulnerability and advisory for CVE-2025-49706. Organizations are encouraged to review all articles and security updates published by Microsoft on July 8, 2025, relevant to the SharePoint platform deployed in their environment.
  • Monitor for POSTs to /_layouts/15/ToolPane.aspx?DisplayMode=Edit
  • Conduct scanning for IPs 107.191.58[.]76, 104.238.159[.]149, and 96.9.125[.]147, particularly between July 18-19, 2025.
  • Update intrusion prevention system and web-application firewall rules to block exploit patterns and anomalous behavior. For more information, see CISA’s Guidance on SIEM and SOAR Implementation.
  • Implement comprehensive logging to identify exploitation activity. For more information, see CISA’s Best Practices for Event Logging and Threat Detection.
  • Audit and minimize layout and admin privileges.

For more information on this vulnerability, please see Eye Security’s reporting and Palo Alto Unit42’s post. CVE-2025-53770 has been added to CISA’s Known Exploited Vulnerabilities (KEV) catalog on July 20, 2025. 

Note: This Alert may be updated to reflect new guidance issued by CISA or other parties.

Organizations should report incidents and anomalous activity to CISA’s 24/7 Operations Center at Report@cisa.gov or (888) 282-0870.  

Disclaimer:  

The information in this report is being provided “as is” for informational purposes only. CISA does not endorse any commercial entity, product, company, or service, including any entities, products, or services linked within this document. Any reference to specific commercial entities, products, processes, or services by service mark, trademark, manufacturer, or otherwise, does not constitute or imply endorsement, recommendation, or favoring by CISA. 

Leviton AcquiSuite and Energy Monitoring Hub

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Leviton
  • Equipment: AcquiSuite, Energy Monitoring Hub
  • Vulnerability: Cross-site Scripting

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to craft a malicious payload in URL parameters that would execute in a client browser when accessed by a user, steal session tokens, and control the service.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of Leviton AcquiSuite and Leviton Energy Monitoring Hub are affected:

  • AcquiSuite: Version A8810
  • Energy Monitoring Hub: Version A8812

3.2 VULNERABILITY OVERVIEW

3.2.1 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION (‘CROSS-SITE SCRIPTING’) CWE-79

The affected products are susceptible to a cross-site scripting (XSS) vulnerability, allowing an attacker to craft a malicious payload in URL parameters, which would execute in a client browser when accessed by a user, steal session tokens, and control the service.

CVE-2025-6185 has been assigned to this vulnerability. A CVSS v3.1 base score of 9.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:N).

A CVSS v4 score has also been calculated for CVE-2025-6185. A base score of 8.7 has been calculated; the CVSS vector string is (AV:N/AC:L/AT:P/PR:N/UI:A/VC:H/VI:H/VA:N/SC:H/SI:H/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Communications
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

notnotnotveg (notnotnotveg@gmail.com) reported this vulnerability to CISA.

4. MITIGATIONS

Leviton has not responded to requests to work with CISA in mitigating this vulnerability. Users of these affected products are welcome to contact Leviton’s customer support for additional information.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities of their own and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time.

5. UPDATE HISTORY

  • July 17, 2025: Initial Publication

Panoramic Corporation Digital Imaging Software

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.5
  • ATTENTION: Low attack complexity
  • Vendor: Panoramic Corporation
  • Equipment: Digital Imaging Software
  • Vulnerability: Uncontrolled Search Path Element

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow a standard user to obtain NT Authority/SYSTEM privileges.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Panoramic Corporation products are affected:

  • Digital Imaging Software: Version 9.1.2.7600

3.2 VULNERABILITY OVERVIEW

3.2.1 UNCONTROLLED SEARCH PATH ELEMENT CWE-427

The affected product is vulnerable to DLL hijacking, which may allow an attacker to obtain NT Authority/SYSTEM as a standard user.

CVE-2024-22774 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2024-22774. A base score of 8.5 has been calculated; the CVSS vector string is (AV:L/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: North America
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Damian Semon Jr. of Blue Team Alpha LLC reported this vulnerability to CISA.

4. MITIGATIONS

The affected software is vulnerable due to an SDK component owned by Oy Ajat Ltd, which is no longer supported. Panoramic Corporation is not the owner of this vulnerable component. Panoramic Corporation did not recommend any specific mitigation for this vulnerability. Users should contact Panoramic Corporation’s support address for further information.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • July 17, 2025: Initial Publication